Sysmex very frequently receives requests for reference ranges for the CBC parameters.

Of course we are glad to provide you with a selection of literature, but reference ranges from different studies are only conditionally transferable, depending on age, gender, ethnicity, etc.

For this reason, the individuals who served as the basis for the selected reference range should be as similar as possible to the patient population to whom the reference range is to be applied.

Currently, there are several scientific publications that provide information about the reference ranges for Sysmex analysers (see table below). These studies were performed mainly using X-Class haematology analysers and it can be seen that the reference ranges differ according to the populations involved. Therefore, published reference ranges from older analysers may be used on newer analysers, but only after validation. Since – as explained above – there are differences between haematological reference ranges from different populations, the information in the table may facilitate the selection of an appropriate reference range as an initial point for validation.

Examples of reference ranges for Sysmex analysers

First author

Citation

Analyser

Parameters

Population

Bohn MK et al.

Int J Lab Hematol XN-3000 CBC+DIFF Canadian, paediatric

Ianni B et al.

Arch Pathol Lab Med XN-1000 CBC+DIFF+RET+PLT-F US American, newborns
Arbiol-Roca A et al. EJIFCC XN-Series CBC+DIFF Spanish
El Graoui et al. Clin Lab. XE-2100 CBC Moroccan
Pekelharing et al. Sysmex J Intl. XE-5000 CBC + DIFF + RET Dutch
Zierk J et al. Clin Chem XE-2100 CBC German, paediatric

Validation of Reference Ranges

Adopting a published reference range without validation is not permissible and can even be dangerous for patients. Its suitability as a reference must be examined in every case. To do this, the IFCC suggests the following procedure:

20 local reference samples are taken and compared with the reference range published elsewhere. If not more than 2 of the 20 samples are outside this range, it can be used.  If 3 - 4 samples are outside the reference range, a further 20 new samples must be taken. If not more than 2 of these 20 samples are outside the range, it can be used. If in the initial verification, 5 or more samples, or more than 2 of a repeated set of samples, are outside the published range, it is not suitable for use as a reference for local patients. The alternative is then a new survey or validation of another, different reference range.

Copyright © Sysmex Europe GmbH. All rights reserved.
Customize your experience

We use cookies to enable you to optimally use our Website and to improve our communication with you. We shall take your selection into account and use only the data for which you give us your consent.

* May lead to restrictions in content and in the user experience
Detail about cookies
Essential cookies
These cookies help to make our website usable by enabling basic features such as page navigation and access to secure areas of our website. Our website cannot function properly without these cookies.
Statistics cookies
By collecting information anonymously, these cookies help us to understand how visitors interact with our website. This information enables us to continually improve our platform.
Marketing cookies
are used to track visitors on websites. The intention is to show advertisements that are relevant and appealing to the individual user and are, therefore, valuable to publishers and third-party advertisers.